YSAV600 Varies Ventilation Modes Anesthetic Machine
-- CE approved portable anesthesia machine
-- Compact light-weight and portable, could be transported by pull-rod case, put on table and assembly on the trolley.
-- Suitable for small animals, Jackson or Bains Absorber is available
-- Specifically designed to accomodate low flow
-- Selectatec-bar and quick change vaporizer mounting device
-- Professional airtight breathing circuit design, provide the stable gas anesthesia, save anesthesia gas consumption, to ensure a clean operating room and laboratory environment.
-- External and reusable soda lime canister, easily watch and replace soda lime.
-- With Oxygen flush function to ensure that clinical anesthesia demand and Oxgyen supply demand.
-- With APL pressure reducing valve to prevent pressure to hurt the animal, ensure the animal safety.
-- Wastewater discharge system
-- Visible inspiration and expiration valve
-- Could upgrade to Oxygen&N2O or Oxygen&Air flow meter freely
-- Module rack with polyester medical material with superior spray technology, high strength and light weight.
-- With transport case, convenient to take and transport
-- Easy to disassemble and assemble all the components, convenient for clean and maintenance.
Specification:
Application: 0.5~100Gkg animal
Ventilation mode: Closed, half-closed, half-open
Tidal volume( Manual Mode): 10~2000ml
Gas pressure: 0.25~0.65Mpa
Flow meter: Oxygen two tube: one tube:0.1~1,one tube:1~10L/M
APL valve: 0.5~6.0Kpa
OXYGEN Flush: 25~75L/min
I:E: As the doctor operation
BPM: As the doctor operation
Monitoring: Airway pressure, respiratory mode.
Three gauges: One for Oxygen source pressure
One for oxygen working pressure
One for airway pressure
Vaporizer
With the latest innovative design of delivery system, vaporizers deliver accurate concentration under varying conditions of variable flow rate, pressure, temperature and so on.
Stably deliver agent, automatic temperature-compensating, deliver accurate agent
Sophisticated delivery system,Stable, prompt, accurate
Various kinds of mounting types.
Safety lock device to prevent the anesthetic gas volatile accidentally.
The big view window is convenient to observe drug usage levels.
Specification
1. The high-precision Vaporizer is designed with a stable concentration output, automatic compensation for pressure, temperature and flow rate.
2. Reliable pneumatic transmission system.
3. Economic and proper configuration of technologies, optional as required by the Customer.
4. Modularized design and upgrading. Upgrading is available as required by the Customer.
YSAV601B Specification
Main body |
High-strength engineering plastic rack, light, beautiful and corrosion resistant |
Scope of application |
adult, child |
Gas source |
O2: 0.27~0.55MPa |
Flow meter |
O2: 0.1~2L/min 2.5~10L/min |
Flow rate of rapid oxygen supply |
25~75L/min |
Low oxygen pressure alarm |
There will be sound alarm when the oxygen
pressure < 0.2MPa |
Vaporizer |
It has the function of automatic compensation based on pressure,
temperature,and flow rate.The regulation range of evaporator
concentration is 0~5 vol%. |
Respiratory circuit |
working mode: all-close, semi-close, semi-open
releasing pressure: 0.5~7 kPa |
Respiratory bellows |
bellows for adults, bellows for children
tidal volume range:0~1500 ml |
Ventilator |
Display mode |
High-definition 4.8" LCD screen display |
Ventilation mode |
IPPV, MANUAL (Manual monitoring of tidal
volume, ventilation volume, respiratory rate) |
Ventilation parameters |
Tidal volume 50~1500ml
Rate 2~99 bpm
SIMV rate 2~20 bpm
I:E 2:1~1:8
Inspiratory trigger pressure -1.0~2.0 kPa
Pressure range 1.0~6.0 kPa |
Parameters for ventilation monitoring |
tidal volume, ventilation volume, IPPV rate,
I/E, peak pressure of airway, inspiratory
trigger pressure, pressure - time waveform |
Security Alarm System |
Airway pressure alarm |
upper limit setting range 1.0~6.0 kPa
low limit setting range 0.4~2.0 kPa |
Per-minute ventilation volume alarm |
upper limit setting range 3.0~30L/min
low limit setting range 1.0~10L/min |
Sustained high-pressure alarm |
it will give alarm when stress have consistently
been higher than 2.5 kPa |
Suffocation alarm |
it will give sound and light alarm if there is
no tidal volume input in 15s |
Power alarm
Power |
AC 220V±10% 50Hz±1Hz |
|
|
|
|
|
Flow range:0.2 to 15L/min
0.2 to 10L/min(normal usage)for concentrations>5 vol%
Filling volume for agent:
About 340ml with dry wick
About 300 ml with moist wick
Mounting type:
selectatec compatible
draeger plug in
cagemount(ISO 23mm tapers)
cagemount with interlock
Net Weight: 26kg
Gross Weight: 30kg
Size:86 X51 X 38CM