Customization: | Available |
---|---|
Type: | Anaesthesia Machine |
Kind: | Gynaecology |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Physical & Environment Specifications
|
|
Dimensions(W×H×D)
|
1403mm×704mm×750mm or 55.2in×27.7in×29.5in
|
Weight
|
100kg or 220.5 lb
|
Storage Humidity:
|
≤93% (noncondensing)
|
Operation Humidity
|
≤80% (noncondensing)
|
Electrical Specifications
|
|
Mains
|
100~240V AC, 50/60Hz
|
Battery
|
24V DC, minimum 120 min
|
Cylinder yokes
|
Optional: O2, N2O, AIR
|
Pneumatic Specifications
|
|
Gas Supply
|
Air, O2, N2O; 280kPa ~ 600kPa
|
Range of fresh gas flow indicators
|
O2, N2O (0.05~10.0 l/min), AIR (0.05~12.0 l/min)
|
O2 flush
|
25~75L/min
|
ACGO
|
Optional auxiliary common gas outlet
|
AGSS
|
Optional anesthetic gas scavenging system
|
Breathing System Specifications
|
|
Operational Modes
|
Closed, Semi-closed, Semi-open
|
CO2 Canister
|
2 Loose fill (3L) or 2 prepacks, optional single canister (1.8 L) |
APL Range
|
Spontaneous breathing (SP) -70 cmH2O
|
Material
|
Autoclavable (except O2 cell and mechanical pressure gauge)
|
CO2 bypas
|
Optional
|
Ventilator Specifications
|
|
Driven Mode
|
Pneumatically driven and electronically controlled
|
Patient Types
|
Adult, Pediatric
|
Ventilation Modes
|
IPPV, PCV, SIMV+PS, PS and Manual
|
Setting Modes
|
Touch key & navigator wheel knob
|
Tidal Volume
|
20~1500ml
|
Ventilation Frequency
|
4~100bpm
|
I:E Ratio
|
4:1~1:8
|
Pressure (inspired) Range
|
5~70cmH2O
|
Pressure (support) Range
|
3~50cmH2O
|
Pressure limit (Pmax)
|
10~70 cm H2O
|
Tslope
|
0~2s
|
Inspiratory Plateau
|
Off, 5%~60%Ti
|
Trigger Sensitivity
|
1~15L/min
|
Compensation
|
Compliance and Leak compensation, fresh gas compensation, altitude compensation
|
Ventilator Monitoring
|
Continuous monitoring of inspiratory O2 concentration, breathing frequency, tidal volume, minute volume, peak airway pressure, PEEP, mean or plateau pressure, CO2 concentration (optional), anesthetic gas concentration (optional) paramagnetic oxygen sensors (optional)
|
Display
|
12.1 in; TFT color touch screen
|
Pressure
|
Ppeak, Pplat,Pmean,PEEP
|
Volumes
|
MV, Vt, Freq
|
Gas
|
O2 Concentration, optional CO2, Agent and MAC
|
Graph Display
|
Waveform of P-t, F-t, V-t, CO2-t (option), Paw-V Loop, V-Flow Loop
|
Alarm
|
MV high/low limit, FiO2 high/low limit, Paw high/low limit, Power failure Breath Rate high limit Subatmospheric pressure Sustained airway pressure Apnea alarm, Alarm Silence(120s) |
Vaporizer
|
|
Support two vaporizers
|
|
Agent
|
Halothane, Enflurane, Isoflurane, Sevoflurane
|
Installation Mode
|
Selectatec® with interlock, optional standby vaporizer parking holder |
Filling Type
|
Pour Fill, Key Fill, Quik-Fil
|
one-stop shopping for operation room | anesthesia machine |
ventilator | |
operating lamp | |
operation table | |
Pendant | |
medical drill and saw | |
ECG | |
patient monitor | |
Defibrillator | |
infusion pump | |
Electrocoagulation | |
oxygen generator | |
electric suction device | |
autoclave sterilizer |