Customization: | Available |
---|---|
Type: | Anaesthesia Machine |
Kind: | Gynaecology |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Main body | High-strength engineering plastic rack, light, beautiful and corrosion resistant |
|
Scope of application | adult, child | |
Gas source | O2: 0.27~0.55MPa | N2O: 0.27~0.55MPa |
Flow meter | O2 : 0.1~1.0L/min | 1.1~10L/min |
N20: 0.1~1.0L/min | 1.1~10L/min | |
Air : 0.1~10L/min | ||
O2 N2O linkage and N2O Stopper | When using nitrous oxide, oxygen concentration>25%; | |
When the oxygen pressure<0.2Kpa, the flow of nitrous oxide would be cut off. |
||
Flow rate of rapid oxygen supply | 25~75L/min | |
Low oxygen pressure alarm | There will be sound alarm when the oxygen pressure < 0.2MPa | |
Vaporizer | It has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5 vol%. Among Halothane,Enflurane, Isoflurane and Sevoflurane, two can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available. |
|
Respiratory circuit | working mode: all-close, semi-close, semi-open APL: 0.5~7 kPa | |
Respiratory Bellows | bellows for adults, tidal volume range:0~1500 ml | |
bellows for children, tidal volume range:0~300 ml | ||
anyone can be chosen for application as | ||
required by the customer | ||
Display mode | High-definition 8.4" TFT color LCD screen display | |
Ventilator | ||
Ventilation mode | IPPV, SIPPV, SIMV, PEEP, MANUAL,SIGH(IMV, VCV, PCV) | |
Ventilation parameters | Tidal volume | 50~1500 ml (20~1500 ml) |
Rate SIMV rate |
1~99 bpm 2~20 bpm |
|
I:E | 2:1~1:8(4:1~1:8) | |
Inspiratory trigger pressure | -1.0~2.0 kPa | |
PEEP | 0~2.0 kPa | |
Pressure range | 1.0~6.0 kPa | |
SIGH | 1.5 times the inspiratory time (70-150 adjustable) | |
Inspiratory platform | 0~50% | |
Parameters for ventilation monitoring | tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, average pressure, pressure - time waveform, flow rate - time waveform,PEEP, inspiratory trigger pressure, Inspiratory platform (autonomous inspiratory rate, lung compliance, airway resistance,flow-volume loop, pressure-volume loop) |
|
Monitoring of oxygen concentration | 21%~100% | |
Security Alarm System | ||
Oxygen Concentration alarm | upper limit setting range 21%~100% | |
low limit setting range 10%~80% | ||
Airway pressure alarm | upper limit setting range 0.3~6.0 kPa | |
low limit setting range 0.2~5.0 kPa | ||
Per-minute ventilation volume alarm | upper limit setting range 3.0~30L/min | |
low limit setting range 1.0~10L/min | ||
Sustained high-pressure alarm | it will give alarm when stress have consistently | |
been higher than 2.5 kPa | ||
Suffocation alarm | it will give sound and light alarm if there | |
is no tidal volume input for 15 seconds | ||
Power alarm | ||
Power(Mechanical Arms) | AC 220V±10% 50Hz±1Hz(UPS, Storage battery) | |
(End-expiratory CO2 monitoring) | ||
Items within brackets are optional |
one-stop shopping for operation room | anesthesia machine |
ventilator | |
operating lamp | |
operation table | |
Pendant | |
medical drill and saw | |
ECG | |
patient monitor | |
Defibrillator | |
infusion pump | |
Electrocoagulation | |
oxygen generator | |
electric suction device | |
autoclave sterilizer |