Customization: | Available |
---|---|
Type: | Anaesthesia Machine |
Kind: | Gynaecology |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Features
1. The high-precision Vaporizer is designed with a stable concentration output, automatic compensation
for pressure, temperature and flow rate. And it is available to choose one kind of gas from Halothan,
Enfluran, Isofluran, and Sevofluran, ensuring a successful anesthesia operation.
2. Reliable pneumatic transmission system.
3. Economic and proper configuration of technologies, optional as required by the Customer.
4. Modularized design and upgrading. Upgrading is available as required by the Customer.
Specifications
1. Anesthesia part
1.Main bodyHigh-strength engineering plastic rack, light,
beautiful and corrosion resistant
2.Scope of applicationAdults & children
3.Gas sourceO2, N2O: 0.27~0.55MPa
4.Flow meterO2: 0.1~1.0L/min 1.1~10L/min
5.Flow rate of rapid oxygen supply 25~75L/min
6.Low oxygen pressure alarm
There will be sound alarm when the oxygen
pressure < 0.2MPa
7.VaporizerIt has the function of automatic compensation based onpressure,temperature,and flow rate.
The regulation range ofevaporator concentration is 0~5 vol%. Customer can choose one kind of gas
container from Halothan, Enfluran, Isofluran and Sevofluran.
8.Respiratory circuit
working mode: all-close, semi-close, semi-open
releasing pressure: 0.5~7 kPa
9.Respiratory bellowstidal volume range:0~1500 ml
2. Ventilator part
1.Display modeHigh-definition 4.8" LCD screen display
2.Ventilation mode IPPV, MANUAL (Manual monitoring of tidal volume, ventilation volume, respiratory
rate)
3.Ventilation parameterTidal volume: 50~1500ml
Rate: 4~80 bpm
I:E: 2:1~1:4
Pressure range: 1.0~6.0 kPa
4.Parameters for ventilation monitoringtidal volume, ventilation volume, IPPV rate, I/E, peak pressure of airway,inspiratory trigger pressure, pressure - time waveform
5.Security Alarm System
Airway pressure alarm: upper limit setting range 1.0~6.0 kPa
low limit setting range 0.4~2.0 kPa
TV alarm:upper limit setting range 0.5~2.0L
low limit setting range 0.02~1.0L
Sustained high-pressure alarm: It will give alarm when stress have consistently been higher than 2.5 kPa
Suffocation alarm: It will give sound and light alarm if there is no tidal volume input in 15s
Power alarm
6.PowerAC 220V±10% 50Hz±1Hz
Usage | For adult and children |
Main body | High-strength engineering plastic rack, light, beautiful and corrosion resistant |
Gas source | O2: 0.27~0.55MPa |
Flow meter | O2: 0.1~2L/min 2.5~10L/min |
Flow rate of rapid oxygen supply | 25~75L/min |
Low oxygen pressure alarm | There will be sound alarm when the oxygen pressure < 0.2MPa |
Vaporizer | It has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5 vol%. Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer. |
Respiratory circuit | working mode: all-close, semi-close, semi-open releasing pressure: 0.5~7 kPa |
Respiratory bellows | bellows for adults, bellows for children tidal volume range: 0~1500 ml |
Ventilator | |
Display mode | High-definition 4.8" LCD screen display |
Ventilation mode | IPPV, MANUAL (Manual monitoring of tidal volume, ventilation volume, respiratory rate) |
Ventilation parameters | Tidal volume 50~1500ml Rate 2~99 bpm SIMV rate 2~20 bpm I: E 2: 1~1: 8 Inspiratory trigger pressure -1.0~2.0 kPa Pressure range 1.0~6.0 kPa |
Parameters for ventilation monitoring | tidal volume, ventilation volume, IPPV rate, I/E, peak pressure of airway, inspiratory trigger pressure, pressure - time waveform |
Security Alarm System | |
Airway pressure alarm | upper limit setting range 1.0~6.0 kPa low limit setting range 0.4~2.0 kPa |
Per-minute ventilation volume alarm | upper limit setting range 3.0~30L/min low limit setting range 1.0~10L/min |
Sustained high-pressure alarm | it will give alarm when stress have consistently been higher than 2.5 kPa |
Suffocation alarm | it will give sound and light alarm if there is no tidal volume input in 15s |
Power alarm Power |
AC 220V± 10% 50Hz± 1Hz |