Display mode |
High-definition 10.4" TFT color LCD screen display |
Ventilation mode |
IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH (IMV, VCV, PCV) |
Ventilation parameters |
Tidal volume |
50~1500 ml (20~1500 ml) |
Rate |
1~99 bpm |
SIMV rate |
1~20 bpm |
I:E |
2:1~1:8 (4:1~1:8) |
Inspiratory trigger pressure |
-1.0~2.0Kpa |
PEEP |
0~2.0Kpa |
Pressure range |
1.0~6.0Kpa |
SIGH |
1.5 times the inspiratory time(70-150 adjustable) |
Inspiratory platform |
0~50% |
Parameters for ventilation monitoring |
tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, average pressure, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure, Inspiratory platform (autonomous inspiratory rate, lung compliance, airway resistance, flow -volume loop, pressure-volume loop) |
Monitoring of oxygen concentration |
21%~100% |
Security Alarm System |
Oxygen Concentration alarm |
upper limit setting range 21%~100% |
low limit setting range 10%~80% |
Airway pressure alarm |
upper limit setting range 0.3~6.0 kPa |
low limit setting range 0.2~5.0 kPa |
Per-minute ventilation volume alarm |
upper limit setting range 3.0~30L/min |
low limit setting range 1.0~10L/min |
Sustained high-pressure alarm |
it will give alarm when stress have consistently been higher than 2.5 kPa |
Suffocation alarm |
it will give sound and light alarm if there is no tidal volume input for 15 seconds |
Power alarm |
Power |
AC 220V±10% 50Hz±1Hz |
|
UPS, Storage battery |
Mechanical Arms |
(Air compressor) |
Items within brackets are optional. |