Customization: | Available |
---|---|
Classification: | Physiological Functions of Diagnosis and Monitoring Equipment |
Type: | ECG |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Product Description | |
YSECG-012A Digital ECG machine 12 Channel 12 leads | |
Performance | |
Input circuit | Isolated input circuit |
Print Format | Automatic/Manual: 6 channel, 12 channel Neutral value+Rhythm lead |
Operation mode | Auto: Automatic lead change: automatic measurement and interpretation MAN: Manual lead change |
Frequency response | 0.05Hz ~ 100Hz |
CMRR | >60dB, >100dB(AC filter in use) |
Time constant | ≥ 3.2s |
Patient leak current | <10µA |
Calibration voltage | 1mV±5% |
Sensitivity | 2.5, 5, 10, 20 mm/mV ±3% |
Noise leve | <15µVp-p |
Input impedance | >50MΩ |
Recording mode | thermal printing system |
Thermal printer | 8 dot/mm(Vertical) 32 dot/mm(Horizontal, 25mm/s) |
Measurement parameters | Ventricle heart rate, P-R interval, QRS Duration, QT/QTC interval, P/QRS/T axis , RV5, SV1, RV5/SV1 amplitude |
Filter | AC filter,EMC filter,DFT filter |
Paper size | 210mm(W)×27m(L) / 210mm(W)×20m(L) roll type |
Skin voltageTolerance | ±500mV |
Papar speed | 6.25, 12.5, 25, 50mm/s |