Customization: | Available |
---|---|
Power Requirements: | AC100-240V, 50/60Hz |
Display: | 12.1 Inch (1024×768 Pixel) Color TFT -LCD |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Physical Specifications | ||||
Dimension and Weight | Dimension | 36cm (W)*23cm (D)*12cm (H) | ||
Weight | 5.3kg(excluding accessories) | |||
Operation Environment | Power requirements | AC100-240V , 50/60Hz | ||
Power | 100VA | |||
Temperature | 0-40ºC | |||
Humidity | 0-93% non-condensing | |||
Fetal/Maternal Performance Specifications | ||||
Performance Specifications |
Display | 12.1 inch (1024×768 pixel) color TFT -LCD | ||
Indicator | Visual on-display and integrated, alarm light, audible alarm |
|||
Battery | Internal rechargeable Li-ion battery | |||
Alarm | Visual on-display and integrated, alarm light, audible alarm |
|||
Memory | 24-hours memory for fetal traces 150 groups patient data |
|||
Printer | America standard/International standard Built-in thermal printer (110mm, 150mm for option) 1/2/3cm/min real-time printing speed, fast print speed (stored traces) |
|||
Paper | Z-fold, thermosensitive | |||
Record Message | ID, name, age, date, time, print speed etc |
|||
CMS | Support connecting to central monitoring station through LAN (TCP/IP) or wireless connection |
|||
Fetal/Maternal Technical Specifications | ||||
Ultrasound (FHR) | Measurement range | US | 30 to 240 bpm | |
Resolution | Display | 1bpm | ||
Printer | 1bpm | |||
jitter@200 bpm | ≤3bpm | |||
Display update rate | 0.25second | |||
Beat to bea change(max.) for ultrasound |
28bpm | |||
US frequency | 1MHz±5%Hz | |||
TOCO | Measurement method | Strain gauge sensor element | ||
Sensitivity | 1 unit =2.5g | |||
Resolution | 1 unit | |||
Measurement range | 500 units | |||
Baseline setting | 0, 5,10, 15, 20 can be selective | |||
Update rate | 0.25 second | |||
Auto offset correction | 3 second after connecting the transducer |
|||
Auto zero adjust |
TOCO value is set to zero negative following a negative Measure value for 5 seconds |
|||
Auto Fetal Movement (AFM) | Technique | Pulsed doppler ultrasound | ||
Range | 0-100% | |||
Resolution | 1% | |||
Automatic fetal movement actograph | ||||
Manual fetal movement marker | ||||
Maternal NIBP(MNIBP) | Operation mode | Manual/automatic/ STAT | |
Measurement unit | mmHg / kPa selectable | ||
Test interval | 0-480 minutes | ||
Accuracy | Max.Std.deviation:8 mmHg(1.1kPa) | ||
Max. mean Error: ±5mmHg(±0.7kPa) | |||
Measurement range | Systolic | 30 to 270 mmHg(4 to 36 kPa) | |
Diastolic | 10 to 245 mmHg(1.5 to 32kPa) | ||
Mean | 20 to 255 mmHg(2.5 to 34kPa) | ||
Pulse rate range | 20~250bpm | ||
Over-pressure protection | Yes | ||
Maternal ECG (MECG) | Heart rate range | 0~300bpm | |
Resolution | 1bpm | ||
Accuracy | ±1bpm/ ±1% | ||
Leads | I, II, III, avL, avR, avF, V | ||
Sweep speed | 12.5mm/s, 25mm/s, 50mm/s | ||
Gain selection | x 0.25, x0.5 ,x1, x2 | ||
Operation mode | diagnostic, monitor, surgery | ||
Maternal SpO2 (MSpO2) | Measurement range | 0%-100% | |
Resolution | 1% | ||
Accuracy | ±2% | ||
Pulse rate range accuracy | ±3bpm | ||
Maternal Temperature (MTEMP) | Technique | Thermistor probe CYSI 400 series compatible |
|
Measurement range | 0ºC-50ºC | ||
Accuracy | ±0.1ºC |