Product Specifications
Syringe Size | 1 ml -20 ml single-use syringe |
Infusion Rate | 1 mm/hr-99 mm/hr |
Alarms | Battery Depleted, Mechanical Error, Occlusion Syringe Disengaged, Completion |
Alerts | Battery Low, User Prompt, Syringe Not Installed, Near End |
Accuracy | ±2% |
Bolus Volume | No limit, min. 0.1 ml |
PCA Function | When the flow rate number is lower than 1/2 of the syringe size, P.C.A. Function can be enabled. Each bolus is 1/4 of current hourly volume, with 30 min locking interval |
OLED Display | Syringe size; rate is set by mm/hr and displayed in ml/hr simultaneously; alarm messages, running status. |
"LOW BATT" LED Indicator | Flashing yellow/red indicator when battery is low or depleted |
Maximum Infusion Pressure | ≤0.12 MPa (with 10 ml syringe) |
Occlusion Alarm Pressure | 0.12 MPa (with 10 ml syringe) CE 0197, IEC 60601 |
Safety Approvals | CE 0197, IEC 60601-1-8:2003, Type CF, IPX1 |
Battery Size | AAx3 |
Dimensions | 166(W)x60(D)x30(H) mm |
Weight | ≤180g (battery weight not included) |
YSZS-AJ5805 portable syringe pump is accurate, reliable infusion solution for Thalassemia(desferral) treatment, pain management, neonatal care and many other ambulatory treatment with low rate small volume. It's compatible with any brand syringe from 1 to 20ml and is programmable in mm/h. The pump can also automatically convert &display the ml/hr rate based on mm/h rate of the calibrated syringe.
Applications
Thalassemia
Parkinson (Apokinon) etc.
Neonatal Care
Immunity Deficiency (Globulin)
Other Applications
Chemotherapy
Pain Management
Epidural Treatment
Safety & Accessories
Alarm Indicators
An immediate identification of pump's alarm and alert messages.
Bolus Cable (PCA)
Patient can push Bolus button to add extra dos-age for quick relief of pain. (Maximum 1 push every 30 minutes for safety).
Bright OLED Screen
Bright & Engergy saving OLED screen displays both infusion rate and alarm message.
Carrying Bag
The carrying bag comes with programmable lock and exit for the PCA cable and IV line for patient.
Syringe Cover
Protects the syringe during the infusion to guarantee the safe & accurate infusion.
Package Contents
The complete package includes a carrying bag with programmable lock for ambulatory use, syringe covers, operating instructions and a set of batteries.
Ease Of Use
Lightweight and compact.
Digital programming with a bright OLED screen.
Clear display of current infusion rate and alarm message.
Powered by 3×AA batteries.